FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2241801 · Received September 9, 2011

Report

Report Number
1423500-2011-11983
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 22, 2011
Report Date
August 22, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE AND THE LOT NUMBER WERE NOT AVAILABLE, THEREFORE, NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. THIS COMPLAINT WAS NOT CONFIRMED. THE ROOT CAUSE WAS UNDETERMINED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING THE HOME CHOICE (HC) ALARM SITUATION CHECK PATIENT LINE (WITH AIR IN LINE), THIS SITUATION OCCURRED DURING INITIAL DRAIN. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HOME PATIENT (HP) AS THE CAREGIVER (CG) STATED THEY HAD AIR IN THE PATIENT LINE. THE TSR ADVISED THE CG WOULD NEED TO START OVER WITH NEW SUPPLIES. THE TSR ADVISED THE CG TO CHECK PATIENT LINE PRIOR TO CONNECTING HP TO MAKE SURE PATIENT LINE PRIMES TO TOP. THE PATIENT WAS INVOLVED, BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. BAXTER CONTACTED THE NURSE ON (B)(6) 2011. THE NURSE STATED THAT THE EVENT WAS DUE TO THE PATIENT AS THE PATIENT WAS NOT ADAPTING WELL TO THERAPY. THE PATIENT HAD CONTINUED ON WITH THERAPY OK AND THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 62 YR HOME CHOICE