COLLEAGUE
Report
- Report Number
- 6000001-2011-23633
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- August 31, 2011
- Report Date
- September 1, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH AN 804:26 FAILURE CODE WAS CONFIRMED DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE OF THE REPORTED PROBLEM WAS A COMMUNICATION ERROR AND SOFTWARE FAILURE. THE MANUAL TUBE RELEASE WAS RESET AND NO FURTHER ACTION WAS REQUIRED. ADDITIONAL INFORMATION: THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.
(B)(4). SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE PUMP BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH AN 804:26 FAILURE CODE. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. ACCORDING TO THE FACILITY REPRESENTATIVE, THERE WAS NO REPORT OF PATIENT INVOLVEMENT, PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |