- 6000001-2011-23623
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- March 7, 2011
- Report Date
- July 21, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE DEPLETED MAIN BATTERIES. TO CORRECT THE CONDITION, THE MAIN BATTERIES WERE REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT. ADDED 510K NUMBER K063696.
(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE WAS RETURNED TO BAXTER (B)(4) AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
UPON REVIEW OF A COLLEAGUE INFUSION PUMP'S EVENT HISTORY BY BAXTER (B)(4) PERSONNEL, A BATTERY DEPLETED SET ALARM WAS FOUND TO HAVE INTERRUPTED DELIVERY. IT IS UNKNOWN IN WHICH CARE AREA THIS EVENT OCCURRED. ACCORDING TO THE FACILITY REPRESENTATIVE, THERE WAS NO REPORT OF PATIENT INVOLVEMENT, PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS INVOLVED A COLLEAGUE P1.5 INFUSION PUMP WITH A USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.08.92.