LIBERTY CYCLER SET, DUEL PATIENT CONNECT
Report
- Report Number
- 8030665-2011-00052
- Event Type
- Malfunction
- Date Received
- August 19, 2011
- Date of Event
- December 23, 2010
- Report Date
- August 4, 2011
- Manufacturer
- REYNOSA MFG.
- Product Code
- FKX
- PMA / PMN Number
- K043363
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
DEVICE HISTORY RECORD REVIEW: THE DHR FOR THIS PRODUCT WAS REVIEWED AND THE FOLLOWING WAS FOUND: THE ASSEMBLY OF THE PRODUCT DID NOT HAVE ANY NCR OR DEVIATIONS DOCUMENTED DURING THE MFG PROCESS RELATED TO THE FAILURE MODE. ALL THE APPLICABLE TESTS DURING THE IN PROCESS AND FINAL INSPECTIONS WERE CONDUCTED IN ORDER TO ENSURE PRODUCT INTEGRITY AND DOCUMENTED WITH ACCEPTABLE RESULTS ACCORDING TO APPLICABLE PROCEDURES. SAMPLE ANALYSIS: THE MFG FACILITY PERFORMED A VISUAL ANALYSIS TO THE ACTUAL SAMPLE RECEIVED AND A HOLE WAS FOUND ON THE FILM OF THE CASSETTE. A LEAK TEST WAS THEN PERFORMED BY INTRODUCING COLORED LIQUID INTO THE CASSETTE AND THE LEAK WAS CONFIRMED, THE HOLE ON THE FILM CAUSES THE LEAK. CONCLUSION: THERE IS NO CAUSE IDENTIFIED DURING MFG OF THIS PRODUCT SINCE THE PRODUCT IS 100% INSPECTED FOR LEAKS PRIOR TO PACKAGING. A VISUAL ANALYSIS OF THE CASSETTE WITH THE MICROSCOPE AND NO MALFORMATIONS SUCH AS FLASHES THAT COULD CAUSE FILM DAMAGE WERE FOUND. CAPA (B)(4) WAS OPENED TO ADDRESS CORRECTIVE ACTIONS FOR CASSETTE LEAKS. (B)(4) HAS BEEN OPENED TO REDUCE THE COMPLAINT RATE AS A CONTINUOUS IMPROVEMENT INITIATIVE.
A PERITONEAL DIALYSIS PT REPORTED THAT SHE RECEIVED AN M31 MESSAGE ON THE CYCLER MACHINE AND WHEN SHE OPENED THE CASSETTE DOOR IN THE MORNING SHE NOTICED SOLUTION LEAKING OUT FROM CYCLER. THERE IS NO REPORTED ILL EFFECT TO THE PT AND THE SAMPLE IS AVAILABLE FOR EVAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIBERTY CYCLER SET, DUEL PATIENT CONNECT | PERITONEAL DIALYSIS TUBING SET | FKX | REYNOSA MFG. | NA | 10KR08102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | LIBERTY DIALYSIS MACHINE |