FDA Adverse Event
Malfunction
Summary report: N
CORDLESS DRIVER 3
MDR report key: 2241662
·
Received September 2, 2011
Report
- Report Number
- 1811755-2011-03205
- Event Type
- Malfunction
- Date Received
- September 2, 2011
- Date of Event
- August 3, 2011
- Report Date
- August 4, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HRX
- PMA / PMN Number
- K943323
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS EVALUATED BY THE MANUFACTURER AND THE REAR MOTOR BEARING FAILED. A FAILED BEARING CAN CAUSE THE DEVICE TO HEAT UP DUE TO THE ADDED FRICTION WHEN BEING OPERATED. THE REAR MOTOR BEARING HAS BEEN REPLACED TO CORRECT THE FAILURE OF THE DEVICE. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE WAS OVERHEATING. THE CUSTOMER DID NOT KNOW WHEN THE EVENT OCCURRED OR IF THERE WAS PT INVOLVEMENT.