FDA Adverse Event Malfunction Summary report: N

CORDLESS DRIVER 3

MDR report key: 2241662 · Received September 2, 2011

Report

Report Number
1811755-2011-03205
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
August 3, 2011
Report Date
August 4, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HRX
PMA / PMN Number
K943323
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED BY THE MANUFACTURER AND THE REAR MOTOR BEARING FAILED. A FAILED BEARING CAN CAUSE THE DEVICE TO HEAT UP DUE TO THE ADDED FRICTION WHEN BEING OPERATED. THE REAR MOTOR BEARING HAS BEEN REPLACED TO CORRECT THE FAILURE OF THE DEVICE. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDPIECE WAS OVERHEATING. THE CUSTOMER DID NOT KNOW WHEN THE EVENT OCCURRED OR IF THERE WAS PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORDLESS DRIVER 3 HRX STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK