FDA Adverse Event Malfunction Summary report: N

STRAIGHT MEDIUM ELITE ATTACHMENT

MDR report key: 2241641 · Received September 2, 2011

Report

Report Number
1811755-2011-03198
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
August 10, 2011
Report Date
August 10, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DRILL ATTACHMENT WAS RETURNED TO THE MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. INTERNAL COMPONENTS WERE EVALUATED, WHICH REVEALED GRITTY BEARING AND FOREIGN DEBRIS CONTAMINATION. THE PRESENCE OF DEBRIS CAN LEAD TO INCREASED FRICTION AMONG ROTATING COMPONENTS, RESULTING IN HEAT BEING GENERATED.

Description of Event or Problem · 1

DURING TESTING, IT WAS REPORTED THAT THE ATTACHMENT OVER-HEATED. THERE WAS NO PT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STRAIGHT MEDIUM ELITE ATTACHMENT ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK