FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2241630 · Received September 2, 2011

Report

Report Number
3008642652-2011-00573
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
July 30, 2011
Report Date
September 1, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CAN COMM TIMEOUTS) WAS CONFIRMED. UPON EVAL THE TRUNK CABLE CONNECTOR WAS SEVERED AND WIRES WERE EXPOSED. THE CAUSE FOR THE DAMAGED TRUNK CABLE CONNECTOR CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED ELECTRODE BELT CONNECTOR. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

AFTER REVIEWING DOWNLOAD DATA OF A (B)(6) FEMALE PT, ZOLL CUSTOMER SUPPORT DISCOVERED "CAN COMM TIMEOUT" ERRORS. THE PT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 62 YR