FDA Adverse Event Malfunction Summary report: N

MICRO DRILL

MDR report key: 2241616 · Received September 2, 2011

Report

Report Number
1811755-2011-03268
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
August 11, 2011
Report Date
August 11, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTION REVEALED CORRODED SOCKETS. CORRODED SOCKETS CAN RESULT IN INTERMITTENT CONNECTION BETWEEN THE CONSOLE AND THE HANDPIECE, AS WELL AS HIGHER IMPEDANCE AT GROUND SOCKETS.

Description of Event or Problem · 1

THE MICRO DRILL WAS SENT FOR SERVICE AND IT RAN ON ITS OWN WITHOUT ACTIVATION DURING PERFORMANCE TESTING. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO DRILL ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK