FDA Adverse Event
Malfunction
Summary report: N
TPS MICRO DRIVER
MDR report key: 2241596
·
Received September 2, 2011
Report
- Report Number
- 1811755-2011-03279
- Event Type
- Malfunction
- Date Received
- September 2, 2011
- Date of Event
- June 24, 2011
- Report Date
- August 8, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS EVALUATED BY THE MFR AND THE EVENT AS REPORTED WAS CONFIRMED. THE CONTACT PINS ARE BURNT AND THE WIRES GOING TO THE FLEX STRIP HAS POOR SOLDER CONNECTIONS THE ORANGE WIRE CAME OFF WHEN WAS FLEXED. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD ACTIVATE WHILE IN SAFE MODE. THIS WAS FOUND WHILE THE DEVICE WAS BEING SERVICED AT THE MFR. THERE WAS NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TPS MICRO DRIVER | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |