FDA Adverse Event Malfunction Summary report: N

TPS MICRO DRIVER

MDR report key: 2241596 · Received September 2, 2011

Report

Report Number
1811755-2011-03279
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
June 24, 2011
Report Date
August 8, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED BY THE MFR AND THE EVENT AS REPORTED WAS CONFIRMED. THE CONTACT PINS ARE BURNT AND THE WIRES GOING TO THE FLEX STRIP HAS POOR SOLDER CONNECTIONS THE ORANGE WIRE CAME OFF WHEN WAS FLEXED. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD ACTIVATE WHILE IN SAFE MODE. THIS WAS FOUND WHILE THE DEVICE WAS BEING SERVICED AT THE MFR. THERE WAS NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TPS MICRO DRIVER ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK