FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2241539 · Received August 9, 2011

Report

Report Number
3008642652-2011-00404
Event Type
Malfunction
Date Received
August 9, 2011
Date of Event
July 4, 2011
Report Date
August 8, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SCREECHING NOISE/WON'T POWER UP) HAS BEEN CONFIRMED. THE CAUSE OF THE RESETS HAS BEEN ISOLATED TO AN INTERMITTENT CONNECTION AT PIN 25 OF THE FLASH MEMORY. THE INTERMITTENT CONNECTION RESULTED IN THE FLASH MEMORY BECOMING CORRUPT. THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURE SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE CORRUPT MEMORY. THE PT RECEIVED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

A (B)(6) FEMALE PT'S DAUGHTER CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PT'S DEVICE MADE A SCREECHING SOUND AND THE SCREEN WENT BLANK. THE DEVICE IS NOW UNABLE TO POWER ON. THE PT WAS ISSUED A REPLACEMENT MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 66 YR