LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2011-00404
- Event Type
- Malfunction
- Date Received
- August 9, 2011
- Date of Event
- July 4, 2011
- Report Date
- August 8, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVAL SUMMARY: DEVICE EVAL OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SCREECHING NOISE/WON'T POWER UP) HAS BEEN CONFIRMED. THE CAUSE OF THE RESETS HAS BEEN ISOLATED TO AN INTERMITTENT CONNECTION AT PIN 25 OF THE FLASH MEMORY. THE INTERMITTENT CONNECTION RESULTED IN THE FLASH MEMORY BECOMING CORRUPT. THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURE SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE CORRUPT MEMORY. THE PT RECEIVED A REPLACEMENT MONITOR.
A (B)(6) FEMALE PT'S DAUGHTER CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PT'S DEVICE MADE A SCREECHING SOUND AND THE SCREEN WENT BLANK. THE DEVICE IS NOW UNABLE TO POWER ON. THE PT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |