FDA Adverse Event Malfunction Summary report: N

THERMOPHORE AUTOMATIC MOIST HEAT PACK

MDR report key: 2241538 · Received August 30, 2011

Report

Report Number
1811605-2011-00105
Event Type
Malfunction
Date Received
August 30, 2011
Date of Event
August 1, 2011
Report Date
August 23, 2011
Manufacturer
BATTLE CREEK EQUIPMENT CO
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THE UNIT STOPPED WORKING, AND IT WAS RETURNED TO US UNDER WARRANTY. EXAMINATION OF THE UNIT REVEALED A BREAK IN THE CORD AT THE ENTRANCE TO THE STRAIN RELIEF AREA NEAR THE SWITCH. THE BREAK PREVENTS THE WIRES IN THE CORD FROM MAKING CONTINUOUS CONTACT AND COULD POSSIBLY EXPOSE BARE WIRES WHEN THE CORD IS BENT. WE ALSO OBSERVED THAT THE CORD OF THE UNIT WAS SEVERELY TWISTED, WHICH SUGGESTS IT WAS BEING USED IN AN UNUSUAL MANNER. THE FAILURE MAY BE A RESULT OF MISUSE OR FAILURE TO FOLLOW INSTRUCTIONS ON THE PART OF THE CONSUMER BY SUBJECTING THE CORD TO EXCESSIVE STRESS. ALTHOUGH USERS ARE INSTRUCTED NOT TO BEND OR TWIST THE CORD, WE ARE TAKING APPROPRIATE ACTION TO MAKE THE DESIGN MORE ROBUST BY INITIATING A CAPA PROJECT ((B)(4)) TO PREVENT THIS FAILURE FROM RECURRING.

Description of Event or Problem · 1

IT WAS REPORTED THE UNIT STOPPED WORKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMOPHORE AUTOMATIC MOIST HEAT PACK ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT CO 055

Patients

Seq Age Sex Outcome Treatment
1