THERMOPHORE AUTOMATIC MOIST HEAT PACK
Report
- Report Number
- 1811605-2011-00105
- Event Type
- Malfunction
- Date Received
- August 30, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 23, 2011
- Manufacturer
- BATTLE CREEK EQUIPMENT CO
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
IT WAS REPORTED THE UNIT STOPPED WORKING, AND IT WAS RETURNED TO US UNDER WARRANTY. EXAMINATION OF THE UNIT REVEALED A BREAK IN THE CORD AT THE ENTRANCE TO THE STRAIN RELIEF AREA NEAR THE SWITCH. THE BREAK PREVENTS THE WIRES IN THE CORD FROM MAKING CONTINUOUS CONTACT AND COULD POSSIBLY EXPOSE BARE WIRES WHEN THE CORD IS BENT. WE ALSO OBSERVED THAT THE CORD OF THE UNIT WAS SEVERELY TWISTED, WHICH SUGGESTS IT WAS BEING USED IN AN UNUSUAL MANNER. THE FAILURE MAY BE A RESULT OF MISUSE OR FAILURE TO FOLLOW INSTRUCTIONS ON THE PART OF THE CONSUMER BY SUBJECTING THE CORD TO EXCESSIVE STRESS. ALTHOUGH USERS ARE INSTRUCTED NOT TO BEND OR TWIST THE CORD, WE ARE TAKING APPROPRIATE ACTION TO MAKE THE DESIGN MORE ROBUST BY INITIATING A CAPA PROJECT ((B)(4)) TO PREVENT THIS FAILURE FROM RECURRING.
IT WAS REPORTED THE UNIT STOPPED WORKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE AUTOMATIC MOIST HEAT PACK | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT CO | 055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |