FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2241530 · Received August 10, 2011

Report

Report Number
3008642652-2011-00410
Event Type
Malfunction
Date Received
August 10, 2011
Date of Event
July 12, 2011
Report Date
August 8, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED ELECTRODE BELT CONNECTOR) HAS BEEN CONFIRMED. UPON EVAL THE TRUNK CABLE CONNECTOR WAS BROKEN AND THE ELECTRODE BELT CABLE WAS DAMAGED AT THE DISTRIBUTION NODE (DN). THE DN TO REAR 2 WIRE THERAPY ELECTRODE PORTION OF THE CABLE WAS PULLED FROM THE STRAIN RELIEF AT THE DN. THE CAUSE FOR THE DAMAGED TRUNK CABLE CONNECTOR AND TRUNK CABLE CANNOT BE POSITIVELY DETERMINED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT CABLE AND CONNECTOR. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS ELECTRODE BELT WAS DAMAGED AT THE CONNECTOR AND HIS MONITOR DISPLAYED SVC CODE 204. THE PT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 50 YR