FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2241519 · Received August 10, 2011

Report

Report Number
3008642652-2011-00408
Event Type
Malfunction
Date Received
August 10, 2011
Date of Event
July 13, 2011
Report Date
August 8, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLE, WIRES EXPOSED) HAS BEEN CONFIRMED. UPON EVALUATION THE TRUNK CABLE CONNECTOR WAS COMPLETELY SEVERED FROM THE ELECTRODE BELT. THE CAUSE FOR THE DAMAGED TRUNK CABLE CONNECTOR CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

THE NURSE OF A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT WIRES WERE STICKING OUT OF THE PATIENT'S ELECTRODE BELT. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.