HOMECHOICE PRO
Report
- Report Number
- 1423500-2011-11963
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- August 9, 2011
- Report Date
- August 23, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE ASSIGNABLE CAUSE OF THE IIPV WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MIN DRAIN VOLUME THRESHOLD. DEVICE MET SPECIFICATIONS RELATIVE TO IIPV IN THE LOGS.
DURING A REVIEW OF THE LOGS OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED DURING DRAIN CYCLE 5. THE PATIENT'S ULTRAFILTRATION (UF) READING WAS 1415 ML, INDICATING THE HOME PATIENT (HP) DRAINED 1415 ML MORE THAN THE LARGEST PRESCRIBED FILL VOLUME OF 2100 ML. THIS MEANT THE TOTAL DRAIN VOLUME WAS APPROXIMATELY 3515 ML, WHICH MEETS THE IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |