FDA Adverse Event
Malfunction
Summary report: N
POSEY SITTER II ALARM
MDR report key: 2241313
·
Received September 2, 2011
Report
- Report Number
- 2020362-2011-00294
- Event Type
- Malfunction
- Date Received
- September 2, 2011
- Report Date
- July 18, 2011
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVAL RESULTS: EVAL OF THE RETURNED PRODUCT SHOWS THE ALARM HAS INTERMITTENT POWER; THE LCD TURNS ON AND OFF WHEN TESTED. ALSO, THERE IS NO ALARM SOUND. THE UNIT DOES NOT PASS FUNCTIONAL TEST. (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTED A CONTINUOUS ALARM TONE WHEN WEIGHT IS APPLIED TO THE SENSOR. THE ALARM WAS TESTED WITH A NEW SENSOR AND THE RESULTS REMAIN THE SAME. THERE IS NO VISIBLE DAMAGE REPORTED TO THE OUTSIDE OF THE ALARM. THERE WAS NO PT INCIDENT OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | POSEY SITTER II ALARM | KMI | J. T. POSEY CO. | 8281 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |