FDA Adverse Event Malfunction Summary report: N

THERMAL LIGATING SHEARS

MDR report key: 2241302 · Received August 19, 2011

Report

Report Number
2954339-2011-00021
Event Type
Malfunction
Date Received
August 19, 2011
Date of Event
July 29, 2011
Report Date
August 19, 2011
Manufacturer
STARION INSTRUMENTS
Product Code
GEI
PMA / PMN Number
K062257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ONE TLS3 35C WAS RETURNED, DECONTAMINATED AND INVESTIGATED. A GENERAL INSPECTION OF THE RETURNED TLS3 35C WAS PERFORMED AND THERE WERE NO COSMETIC ISSUES. THE LEFT JAW BOOT WAS BROKEN AT THE TIP, CONFIRMING THE COMPLAINT. THE RIGHT HEAT SPREADER WAS BENT INWARDS. THE BOOT TEAR AND THE HEAT SPREADER BEING BENT INWARDS ARE CONSISTENT WITH THE JAWS NOT BEING COMPLETELY CLOSED WHEN INSERTING INTO THE CANNULA. INCOMPLETE CLOSURE DURING INSERTION OR REMOVAL FROM A CANNULA CAN CAUSE THE JAWS TO BE APART, FORCING THEM TO BE IN CONTACT WITH THE CANNULA; IF THIS OCCURS, THE BOOT OR OTHER COMPONENTS CAN GET DAMAGED. ALTHOUGH DEFINITIVE DAMAGE WAS FOUND, THE ROOT CAUSE COULD NOT BE DETERMINED. THERE HAS BEEN NO ADDITIONAL COMPLAINTS RECEIVED WITH HEAT SPREADERS BENT IN SUCH A MANNER.

Description of Event or Problem · 1

THE SURGEON ACTIVATED THE INSTRUMENT APPROXIMATELY 2 TIMES AND NOTICED A PIECE OF THE TIP BROKE OFF. THE SURGEON RETRIEVED THE PIECE. NO PATIENT INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMAL LIGATING SHEARS TLS3 35C GEI STARION INSTRUMENTS 132-136D 908014

Patients

Seq Age Sex Outcome Treatment
1