COLLEAGUE
Report
- Report Number
- 6000001-2011-23558
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- August 15, 2011
- Report Date
- August 15, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED BY SERVICE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE A DAMAGED POLE CLAMP. TO CORRECT THE CONDITION, THE POLE CLAMP WAS REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SENT.
DURING PRODUCT EVALUATION BY BAXTER PERSONNEL, A COLLEAGUE INFUSION PUMP FAILED A MOUNTING CLAMP CHECK. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. THIS INVOLVED A REMEDIATED COLLEAGUE PUMP WITH USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.09.90. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |