FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2241159 · Received August 17, 2011

Report

Report Number
8020893-2011-00350
Event Type
Malfunction
Date Received
August 17, 2011
Date of Event
August 4, 2011
Report Date
August 4, 2011
Manufacturer
CIVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING WHILE IN USE ON A PT. THE PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE PURITAN BENNETT CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE BDU PCB. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK CIVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1