FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2241154 · Received August 17, 2011

Report

Report Number
8020893-2011-00352
Event Type
Malfunction
Date Received
August 17, 2011
Date of Event
July 21, 2011
Report Date
July 25, 2011
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BG
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR SHUT DOWN WHEN PLACED ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE CSE INSPECTED THE DEVICE AND REPLACED THE POWER SUPPLY. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840