FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2241134 · Received August 17, 2011

Report

Report Number
1720753-2011-21109
Event Type
Malfunction
Date Received
August 17, 2011
Date of Event
August 8, 2011
Report Date
August 17, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REP RESEATED THE CONNECTORS AND ADJUSTED THE CAMERA POWER SUPPLY. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD INTERMITTENTLY PRODUCE FLUOROSCOPIC X-RAY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1