ANIMAS INSULIN INFUSION PUMP
MDR report key: 2240982
·
Received September 9, 2011
Report
- Report Number
- 2531779-2011-06674
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Report Date
- August 11, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT THE OK BUTTON REQUIRED SEVERAL PRESSES TO RESPOND. THE PATIENT REPORTEDLY WORE THE PUMP CLIPPED TO AN UNDERGARMENT AND DID NOT CLEAN THE PUMP.