FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2240949
·
Received August 9, 2011
Report
- Report Number
- 3008642652-2011-00413
- Event Type
- Malfunction
- Date Received
- August 9, 2011
- Date of Event
- June 30, 2011
- Report Date
- August 9, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 204) HAS BEEN CONFIRMED. UPON EVALUATION, THE J702 CONNECTOR ON THE DISTRIBUTION NODE (DN) PCA BOARD WAS BENT WHICH HALTED COMMUNICATION BETWEEN THE DN AND THE MONITOR. THE ROOT CAUSE OF THE BENT J702 CONNECTOR WAS EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE BENT J702 CONNECTOR. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE WAS GETTING A MESSAGE THAT "THE DEVICE WAS NOT WORKING" (SERVICE CODE 204). THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR |