FDA Adverse Event Malfunction Summary report: N

PROXIMATE ILS CURVED CIR STAPL

MDR report key: 2240898 · Received September 9, 2011

Report

Report Number
3005075853-2011-03714
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 18, 2011
Report Date
August 22, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K983536
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION WITHOUT STAPLES PRESENT AND WITH THE BREAKAWAY WASHER UNCUT AND INDENTED, INDICATING THAT THE DEVICE HAD NOT BEEN FIRED THROUGH A FULL FIRING STROKE OR POSSIBLY THAT THE ORANGE INDICATOR WAS NOT FULLY INTO THE SAFE GREEN FIRING RANGE. IT SHOULD BE NOTED THAT BEFORE FIRING THE DEVICE THE ORANGE INDICATOR SHOULD BE FULLY WITHIN THE GREEN RANGE OF THE GAP SETTING SCALE. IN ADDITION IT SHOULD BE NOTED THAT IF THE FIRING SEQUENCE IS NOT COMPLETE (PLASTIC TO PLASTIC) STAPLES COULD BE PARTIALLY DEPLOYED WITHOUT CUTTING THE WASHER. THE DEVICE WAS RELOADED WITH STAPLES AND TESTED FOR FUNCTIONALITY WITH A TEST WASHER; THE DEVICE FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A HARTMANN'S REVERSAL PROCEDURE, THE GUN CUT WITHOUT FORMING STAPLES. "REG" FIRED THE STAPLER AND THE CONSULTANT HAD TO HAND-SEW THE ANASTOMOSIS. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE ILS CURVED CIR STAPL STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1