FDA Adverse Event Malfunction Summary report: N

HOMEPUMP ECLIPSE, 400ML, 100ML/HR

MDR report key: 2240873 · Received August 16, 2011

Report

Report Number
2026095-2011-00214
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
July 12, 2011
Report Date
July 18, 2011
Manufacturer
I-FLOW CORP.
Product Code
MEB
PMA / PMN Number
K052117
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

METHOD: SAMPLE HAS NOT YET BEEN RECEIVED, BUT WAS REPORTED TO BE AVAILABLE. A REVIEW OF THE DEVICE HISTORY RECORDS (DHR) WAS CONDUCTED FOR THE LOT NUMBER AND INCIDENT REPORTED. THE DEVICE PASSED ALL MFG SPECS PRIOR TO RELEASE. INFO PROVIDED INDICATES THE PUMP WAS UNDER-FILLED TO 150ML, WHICH DOES FLOW FASTER THAN A NOMINALLY FILLED PUMP. THE DIRECTIONS FOR USE ((B)(4)) CONTAINS A CAUTION STATEMENT: FILLING THE PUMP LESS THAN NOMINAL RESULTS IN FASTER FLOW RATE. RESULTS: WITHOUT THE ACTUAL PRODUCT, AN ANALYSIS CANNOT BE CONDUCTED. CONCLUSIONS: THE SAMPLE WILL BE EVALUATED WHEN RECEIVED AND A FOLLOW-UP REPORT WILL BE FILED.

Description of Event or Problem · 1

DRUG/DILUENT: LEVAQUIN. FILL VOLUME: 150ML AND FLOW RATE: 100ML/HR. PROCEDURE: INFECTION TREATMENT. CATHPLACE: DOUBLE LUMEN PICK, TYPICALLY IN ARM. DRUG INFUSED IN 55 MINUTES. NO HARM TO PT. PER DFU : NOMINAL FLOW RATE: 100ML/HR. NOMINAL FILL VOLUME: 400ML. MAXIMUM FILL VOLUME: 500ML. THE INFUSION DELIVERY TIME IS APPROX 1:35 HOURS WHEN FILLED TO NOMINAL VOLUME AND FLOW RATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMEPUMP ECLIPSE, 400ML, 100ML/HR ELASTOMERIC PUMP MEB I-FLOW CORP. E401000 142508

Patients

Seq Age Sex Outcome Treatment
1 57 YR