FDA Adverse Event Summary report: N

HAUSTED VIC VIDEOFLUOROSCOPIC IMAGING CHAIR

MDR report key: 2240746 · Received September 2, 2011

Report

Report Number
2240746
Date Received
September 2, 2011
Date of Event
July 1, 2011
Report Date
September 2, 2011
Manufacturer
HAUSTED PATIENT HANDLING SYSTEMS, LLC
Product Code
FZK
Report Source
User Facility report
Reporter Location
WI, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

WHILE SCANNING PATIENT'S ESOPHAGUS USING A VIDEOFLUOROSCOPIC SWALLOW CHAIR, THE FLUOROSCOPY ARM WAS MOVED TOO LOW AND AN ALARM SOUNDED. MOVEMENT FUNCTIONS OF THE FLUORO ARM THEN STOPPED AND DID NOT ALLOW THE ARM TO BE RAISED, TRAPPING THE PATIENT'S LEGS. UPON INVESTIGATION, THERE IS A BUTTON ON THE MAIN CONSOLE AND ON REMOTE CONTROL THAT WILL RESTORE LIFTING CAPABILITY TO THE FLUOROSCOPY ARM, BUT THE STAFF DID NOT RECALL THAT DURING THE EVENT.THERE ARE "HALL EFFECT" SWITCHES IN THE COWLING THAT WERE REPLACED IN 2008 AND KNOWN TO BE FUNCTIONAL AT THE TIME OF THE OCCURRENCE. THESE SENSE AN OBSTRUCTION AND STOP FURTHER MOVEMENT. THESE WERE OPERATIONAL, AS WAS THE MECHANISM TO RESTORE THE LIFTING FUNCTION. IT APPEARS PATIENT'S LOW BMI MAY HAVE CONTRIBUTED TO EVENTS SINCE CHAIR IS LIMITED IN HEIGHT, AND PATIENT WAS LOW IN THE CHAIR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HAUSTED VIC VIDEOFLUOROSCOPIC IMAGING CHAIR CHAIR, IMAGING, VIDEOFLUOROSCOPIC FZK HAUSTED PATIENT HANDLING SYSTEMS, LLC * *
2 MULTIDIAGNOST 4 X-RAY SYSTEM, RAD/FLUORO JAA PHILLIPS MEDICAL SYSTEMS NORTH AMERICA, INC MULTIDIAGNOST 4 *

Patients

Seq Age Sex Outcome Treatment
1 84 YR