75MM SELECTABLE NEW TLC
Report
- Report Number
- 3005075853-2011-03707
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Report Date
- August 19, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K092577
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). SWING TAB IN LOCKED POSITION. THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A CARTRIDGE RELOAD LOADED IN THE INSTRUMENT. THE CARTRIDGE RELOAD HAD THE SWING TAB IN THE LOCKED POSITION, AND THE PROXIMAL ONE DRIVER UP AND THE REMAINING DRIVERS WERE DOWN WITH STAPLES PRESENT. IT COULD NOT BE DETERMINED IF THE REPORTED INCIDENT WAS THE RESULT OF A PREMATURE LOCKOUT, OR THE RESULT OF AN INTERRUPTED FIRING STROKE. THE SWING TAB WAS RESET AND THE CARTRIDGE WAS LOADED INTO THE DEVICE FOR FUNCTIONAL TESTING. THE DEVICE ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER SHAPE. WHILE NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT, IT SHOULD BE NOTED THAT 100% INSPECTION IS PERFORMED BY MEANS OF (B)(4) TO INSURE THE SWING TAB IS PRESENT AND UNLOCKED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE DID NOT STAPLE AT ALL. NO FURTHER INFORMATION IS AVAILABLE. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 75MM SELECTABLE NEW TLC | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | G4TY9Z |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RELOAD-SR75 |