CURVE CUTTER STAPLER
Report
- Report Number
- 3005075853-2011-03709
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 19, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K040038
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT.INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH A CARTRIDGE RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED VOID OF STAPLES, WITH THE WASHER COMPLETELY CUT, WITH THE DRIVERS EXPOSED AND WITH THE KNIFE RECESS BELOW THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH AN ENGINEERING SAMPLE RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE LOCKOUT AND THE RELOAD LOCKOUT WERE FUNCTIONAL. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A COLORECTAL RESECTION PROCEDURE, THE DEVICE WOULD NOT STAPLE BUT CUT. THE DEVICE CUT WELL BUT DID NOT STAPLE. THE DEVICE WAS USED ON A COLON. THE CARTRIDGE USED WAS THE FIRST ONE AND IT WAS A BLUE ONE. THE DEVICE WAS NOT FIRED NEXT TO AN EXISTING STAPLE LINE. NO RESISTANCE WAS FELT BY THE SURGEON. AFTER USE ALL BUTTONS AND TRIGGERS AUTOMATICALLY RETURNED TO THEIR ORIGINAL POSITION. THE SURGEON DID NOT HAVE ANY DIFFICULTY REMOVING THE DEVICE FROM THE PATIENT TISSUE. THE SURGEON HAD TO DO A MANUAL SUTURE TO FINISH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CURVE CUTTER STAPLER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | G4TX60 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARTRIDGE |