FDA Adverse Event Malfunction Summary report: N

CURVE CUTTER STAPLER

MDR report key: 2240675 · Received September 9, 2011

Report

Report Number
3005075853-2011-03709
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 1, 2011
Report Date
August 19, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K040038
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT.INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH A CARTRIDGE RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED VOID OF STAPLES, WITH THE WASHER COMPLETELY CUT, WITH THE DRIVERS EXPOSED AND WITH THE KNIFE RECESS BELOW THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH AN ENGINEERING SAMPLE RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE LOCKOUT AND THE RELOAD LOCKOUT WERE FUNCTIONAL. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A COLORECTAL RESECTION PROCEDURE, THE DEVICE WOULD NOT STAPLE BUT CUT. THE DEVICE CUT WELL BUT DID NOT STAPLE. THE DEVICE WAS USED ON A COLON. THE CARTRIDGE USED WAS THE FIRST ONE AND IT WAS A BLUE ONE. THE DEVICE WAS NOT FIRED NEXT TO AN EXISTING STAPLE LINE. NO RESISTANCE WAS FELT BY THE SURGEON. AFTER USE ALL BUTTONS AND TRIGGERS AUTOMATICALLY RETURNED TO THEIR ORIGINAL POSITION. THE SURGEON DID NOT HAVE ANY DIFFICULTY REMOVING THE DEVICE FROM THE PATIENT TISSUE. THE SURGEON HAD TO DO A MANUAL SUTURE TO FINISH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CURVE CUTTER STAPLER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA G4TX60

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE