FDA Adverse Event
Malfunction
Summary report: N
PRESSUREGUARD EASY AIR
MDR report key: 2240614
·
Received April 12, 2007
Report
- Report Number
- 1041130-2007-00004
- Event Type
- Malfunction
- Date Received
- April 12, 2007
- Date of Event
- March 23, 2007
- Report Date
- April 12, 2007
- Manufacturer
- SPAN-AMERICA MEDICAL SYSTEM, INC.
- Product Code
- FNM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ADDITIONAL INFO: SERIAL #: (B)(4). THIS DEVICE FAILURE IS REPORTED AS A RESULT OF A PREVIOUS, SIMILAR FAILURE ASSOCIATED WITH A PT FALLING OFF THE MATTRESS. WE HAVE RECEIVED 75 COMPLAINTS OF WELD FAILURES IN THE TOP COVER OF THIS DESIGN, WITH ONE FAILURE ASSOCIATED WITH PT INJURY. THIS DESIGN WAS DISCONTINUED IN (B)(6) 2006.
Description of Event or Problem · 1
THIS IS A REPEAT PRODUCT DEFECT OCCURRENCE FOR PREVIOUS MDR 1041130-2007-00001 WITH A REPORT DATE OF (B)(6) 2007, PRESSUREGUARD EASY AIR, AN ALTERNATING PRESSURE MATTRESS, FOR WELD FAILURE IN THE TOP COVER FABRIC. THIS CUSTOMER COMPLAINT IS FOR FAILURE FOR THREE MATTRESSES. THESE EVENTS DID NOT RESULT IN AN ADVERSE EVENT FOR THE PT.