FDA Adverse Event Other Summary report: N

DEXTRUS 4135

MDR report key: 2240589 · Received August 19, 2011

Report

Report Number
1028232-2011-01855
Event Type
Other
Date Received
August 19, 2011
Date of Event
June 23, 2011
Report Date
August 8, 2011
Manufacturer
BIOTRONIK SE & CO., KG.
Product Code
NVZ
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS, THEREFORE, BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MFG PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MFG PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MFG.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFO THAT THIS RIGHT ATRIAL (RA) LEAD HAD DISLODGED. ELEVATED THRESHOLDS WERE NOTED AT A F/U, AND A CHEST X-RAY WAS TAKEN TO CONFIRM THE DISLODGEMENT. NO SURGICAL INTERVENTION HAS YET BEEN TAKEN. NO ADVERSE PT EFFECTS WERE REPORTED. THE LEAD WAS REPROGRAMMED AND REMAINS IN SERVICE. AS NO FURTHER INFO CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFO BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4135 PACER LEAD NVZ BIOTRONIK SE & CO., KG. 358744

Patients

Seq Age Sex Outcome Treatment
1 UNK Other