FDA Adverse Event
Injury
Summary report: N
TENDRIL ST
MDR report key: 2240509
·
Received September 9, 2011
Report
- Report Number
- 2017865-2011-05716
- Event Type
- Injury
- Date Received
- September 9, 2011
- Date of Event
- July 22, 2011
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVN
- PMA / PMN Number
- P960013
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NO MEDWATCH FORM WAS RECEIVED.
Additional Manufacturer Narrative · 1
FINAL ANALYSIS FOUND THAT THE DISTAL INSULATION WAS DAMAGED AT 31.5 CM FROM THE CONNECTOR PIN. THE DAMAGED INSULATION CAUSED A SHORT CIRCUIT, WHICH CAUSED THE LOW LEAD IMPEDANCE PROBLEM.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT EXPERIENCED DIAPHRAGMATIC STIMULATION. THE RIGHT VENTRICULAR LEAD IMPEDANCE WAS FOUND TO OCCASIONALLY DROP BELOW 100 OHMS. THE LEAD WAS REMOVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TENDRIL ST | PERMANENT PACEMAKER ELECTRODE | NVN | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1888TC/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |