FDA Adverse Event Injury Summary report: N

CONFIRM

MDR report key: 2240468 · Received September 9, 2011

Report

Report Number
2017865-2011-05931
Event Type
Injury
Date Received
September 9, 2011
Date of Event
July 27, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
MXC
PMA / PMN Number
K081365
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4) - FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT TWO CIRCUIT BLOCKS WERE SHORTED WHICH PREMATURELY DRAINED THE LOOP RECORDER'S BATTERY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE EXHIBITED PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONFIRM IMPLANTABLE CARDIAC MONITOR MXC ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION DM2100 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention