OT PING METER
Report
- Report Number
- 2939301-2011-08709
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Report Date
- September 1, 2011
- Manufacturer
- LIFESCAN INC.
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 (11/04/2011)-DEVICE EVALUATION: THE METER AND TEST STRIPS INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: ON (B)(6), 2011 THE METER WAS TESTED AND THE PRIMARY COMPLAINT WAS NOT CONFIRMED. HOWEVER, THE METER WAS FOUND TO HAVE CONTAMINATION AROUND SPC PINS AND SPC PIN 2 WAS LIFTED HIGH. ON (B)(6), 2011 THE TEST STRIPS WERE TESTED AND NO FAULTS WERE FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT; 510(K)# IS K082590.
THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE DOES NOT TURN ON WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN INC. | 3025900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |