FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 2240330 · Received September 9, 2011

Report

Report Number
2017865-2011-05678
Event Type
Injury
Date Received
September 9, 2011
Date of Event
July 20, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

FINAL ANALYSIS REVEALED THAT THE DISTAL INSULATION WAS ABRADED OPEN AT 4.8 CM TO 5.5 CM AND AT 3.3 CM TO 4.3 CM FROM THE DISTAL TIP. THE DISTAL INSULATION WAS ABRADED FROM THE INSIDE FROM PRESSURE AND MOVEMENT AGAINST THE COILS. AN ELECTRICAL SHORT WAS NOTED BETWEEN COILS DUE TO OPEN DISTAL INSULATION WHICH RESULTED IN THE REPORTED IMPEDANCE ANOMALY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED A GRADUAL DECLINE IN BIPOLAR IMPEDANCE TO LESS THAN 200 OHMS AND HIGH THRESHOLDS. THE LEAD WAS REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL ST PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1888TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1 60 YR Required Intervention