FDA Adverse Event
Injury
Summary report: N
TENDRIL ST
MDR report key: 2240330
·
Received September 9, 2011
Report
- Report Number
- 2017865-2011-05678
- Event Type
- Injury
- Date Received
- September 9, 2011
- Date of Event
- July 20, 2011
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Additional Manufacturer Narrative · 1
FINAL ANALYSIS REVEALED THAT THE DISTAL INSULATION WAS ABRADED OPEN AT 4.8 CM TO 5.5 CM AND AT 3.3 CM TO 4.3 CM FROM THE DISTAL TIP. THE DISTAL INSULATION WAS ABRADED FROM THE INSIDE FROM PRESSURE AND MOVEMENT AGAINST THE COILS. AN ELECTRICAL SHORT WAS NOTED BETWEEN COILS DUE TO OPEN DISTAL INSULATION WHICH RESULTED IN THE REPORTED IMPEDANCE ANOMALY.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED A GRADUAL DECLINE IN BIPOLAR IMPEDANCE TO LESS THAN 200 OHMS AND HIGH THRESHOLDS. THE LEAD WAS REMOVED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TENDRIL ST | PERMANENT PACEMAKER ELECTRODE | NVN | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1888TC/52 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Required Intervention |