FDA Adverse Event Malfunction Summary report: N

TENDRIL ST

MDR report key: 2240197 · Received September 9, 2011

Report

Report Number
2017865-2011-05642
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
February 21, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
P960013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. RELIABILITY LABORATORY TECHNICIAN, (B)(4). (B)(4) FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND MEDICAL ADHESIVE ON THE RING ELECTRODE, CAUSING THE LEAD TO EXHIBIT HIGH LEAD IMPEDANCE IN THE BI-POLAR CONFIGURATION.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE LEAD EXHIBITED UNACCEPTABLE THRESHOLDS. THE LEAD WAS NOT IMPLANTED AND REPLACED.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL ST PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1888TC/46 NA

Patients

Seq Age Sex Outcome Treatment
1 74 YR