FDA Adverse Event Malfunction Summary report: N

IMMAGE® IMMUNOCHEMISTRY SYSTEM

MDR report key: 2240188 · Received September 9, 2011

Report

Report Number
2050012-2011-05090
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 11, 2011
Report Date
August 11, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JQX
PMA / PMN Number
K962294
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER PERFORMED TROUBLESHOOTING WITH BEC CTS (CUSTOMER TECHNICAL SUPPORT) AND NOTED THAT THE SAMPLE AND REAGENT PROBE WASH STATIONS WERE NOT DRAINING AND OVERFLOWING AS A RESULT OF A LACK OF VACUUM. CTS GENERATED A SERVICE REQUEST. A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE (B)(4) 2011 AND REPLACED A DIAPHRAGM DUE TO THE LOW VACUUM ISSUE. REPAIR WAS VERIFIED PER ESTABLISHED PROCEDURES AND RESULTS MET PUBLISHED PERFORMANCE SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT THE SAMPLES AND REAGENT PROBE WASH STATIONS IN THE IMMAGE IMMUNOCHEMISTRY SYSTEM WERE NOT DRAINING AND WERE OVERFLOWING. NO INJURY OR EXPOSURE WAS REPORTED AND NO ERRONEOUS RESULTS WERE GENERATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMMAGE® IMMUNOCHEMISTRY SYSTEM CLINICAL CHEMISTRY, NEPHELOMETER JQX BECKMAN COULTER INC. IMMAGE NA

Patients

Seq Age Sex Outcome Treatment
1