AUTOSOFT XC
Report
- Report Number
- 3003442380-2025-11295
- Event Type
- Malfunction
- Date Received
- July 4, 2025
- Date of Event
- June 19, 2025
- Report Date
- September 25, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244016620
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4 AND MANUFACTURING DATE UNDER H4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. REVISION 21 OF (B)(4) DOES NOT REQUIRE A COMPLAINT THAT IS TYPE 2 REPORTABLE TO OPEN A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS OPENED AGAINST A PREVIOUS REVISION OF (B)(4). COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010681, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 24-SEP-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL "6010681". THE COUNT OF COMPLAINT IS 0 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DHR REVIEW: THE LOT 6010681 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 120 AND MANUFACTURED IN THE LINE INSET 7 ON 05/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. GLUE-TUBING LOT: THE LOT 4K05979 WAS MANUFACTURED ACCORDING TO THE[?]WI VERSION 12 AND MANUFACTURED IN THE MACHINE IGTL01 ON 13/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED AN INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025. THE BLOCKAGE WAS IN THE TUBING. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2768468 | AUTOSOFT XC | UNO INSET I 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1001680 | 6010681 | 05705244016620 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Female |