FDA Adverse Event Injury Summary report: N

TENDRIL SDX

MDR report key: 2240160 · Received September 9, 2011

Report

Report Number
2017865-2011-05598
Event Type
Injury
Date Received
September 9, 2011
Date of Event
July 5, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
P960013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED A CAPTURE THRESHOLD OF 5.0 V, 0.4 MS AND AN INCREASE IN LEAD IMPEDANCE FROM 500 OHMS TO 1700 OHMS. THE LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL SDX PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1688TC/58 NA

Patients

Seq Age Sex Outcome Treatment
1 90 YR Required Intervention