FULL FACE MASK
Report
- Report Number
- 9611451-2011-00553
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- August 8, 2011
- Report Date
- August 12, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- MNT
- PMA / PMN Number
- K060044
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- RESPIRATORY THERAPIST
Narratives
(B)(4). LOT DATE AND DEVICE MANUFACTURER DATE - 090807 - 08/07/2009; 100618 - 06/18/2010. EIGHTEEN COMPLAINT MASKS WERE RETURNED, 14 HAD LOT 090807 AND 3 HAD LOT 100618. THE RETURNED COMPLAINT MASKS WERE VISUALLY INSPECTED. THE VISUAL INSPECTION REVEALED THAT THE ON 10 OF THE RETURNED COMPLAINT MASKS THE SEAL IS PARTLY SEPARATED FROM THE MASK BASE STARTING NEAR THE CHIN. THE MASK SEAL FOR THE OTHER 3 COMPLAINT MASKS THE MASK SEAL HAS COMPLETELY SEPARATED FROM THE MASK BASE. THE VISUAL INSPECTION OF ALL COMPLAINT MASKS REVEALED THAT THE SEAL WAS GLUED PROPERLY WITH A CONTINUOUS BEAD OF GLUE. INDENTS IN THE GLUE LINE SHOW THAT THE SEAL WAS PROPERLY INSERTED INTO THE BASES. A LOT CHECK REVEALED (B)(4) OTHER COMPLAINT OF THIS NATURE FOR LOT NUMBER 090807 A LOT CHECK REVEALED (B)(4) OTHER COMPLAINT OF THIS NATURE FOR LOT NUMBER 100618 THE RT040 FULL FACE MASK FEATURES A MASK BASE, SEAL AND HEADGEAR. THE MASK SEAL IS INSERTED INTO THE MASK BASE AND IS SECURED WITH GLUE. THE RT040 MASK IS SUBJECTED TO POST-PRODUCTION TESTS TO ENSURE THE SEAL ON THE MASK CAN WITHSTAND A REMOVAL FORCE GREATER THAN 100N. WE ARE UNABLE TO DETERMINE THE ROOT CAUSE OF THE SEPARATION. THE VISUAL INSPECTION OF THE RETURNED DEVICES SHOW THAT THE MASK WAS ASSEMBLED AND GLUED PROPERLY. THIS SUGGESTS THAT THE SEPARATION OCCURRED POST-PRODUCTION, POSSIBLY DUE TO THE DRYING OUT OR SEPARATION OF THE GLUE DUE TO ADVERSE EFFECTS OF TRANSPORT OR STORAGE CONDITIONS. (B)(4).
A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THE MASK SEALS OF THE RT040 FULL FACE MASK SEPARATED FROM THE SHELL CAUSING IT TO LEAK. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FULL FACE MASK | MNT | MNT | FISHER & PAYKEL HEALTHCARE LTD | RT040 | SEE H.10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |