ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-06659
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Report Date
- August 14, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE AND CORROSION WAS OBSERVED UNDER THE BUTTON CONTACTS.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2011, THE PATIENT CONTACTED ANIMAS ALLEGING THAT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE. THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THE PUMP WAS REPLACED. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE PATIENT CLAIMED THAT THE PUMP WAS UNRESPONSIVE TO BUTTON PRESSES. THERE IS NO EVIDENCE HOWEVER, THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY BLOOD GLUCOSE VALUES, SYMPTOMS, OR TREATMENT SUGGESTIVE OF A SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |