FDA Adverse Event Malfunction Summary report: N

BD FACSLYRIC¿ FLOW CYTOMETER

MDR report key: 22399344 · Received July 3, 2025

Report

Report Number
3027394506-2025-00031
Event Type
Malfunction
Date Received
July 3, 2025
Date of Event
June 17, 2025
Report Date
September 8, 2025
Manufacturer
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
Product Code
OYE
UDI-DI
00382906635181
PMA / PMN Number
BK230835
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E.1. INITIAL REPORTER PHONE #: (B)(6). E.1. INITIAL REPORTER FACILITY NAME: (B)(6). G.5. PMA / 510(K) ADDITIONAL #170974; K201814. H.3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6: INVESTIGATION SUMMARY: BASED ON THE INVESTIGATION RESULTS, THE REPORTED ISSUE FOR THE LARGE SAMPLE RESIDUE WAS CONFIRMED. INVESTIGATION RESULTS THAT WERE PERFORMED ON THE INDICATED FAILURE MODE WERE THE FOLLOWING: DEVICE HISTORY REVIEW (DHR), COMPLAINT HISTORY REVIEW, AND SERVICE HISTORY REVIEW. THE POTENTIAL CAUSE FOR THE LARGE SAMPLE RESIDUE WAS FAULTY PUMP AND DEGASSER. THE FIELD SERVICE REPRESENTATIVE (FSR) REPLACED THE FAULTY PARTS. AFTER THE REPLACEMENT, THE INSTRUMENT WAS TESTED AND WAS CONFIRMED TO BE PERFORMING ACCORDING TO SPECIFICATIONS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH BD FACSLYRIC¿ FLOW CYTOMETER CARRYOVER INVOLVING PATIENT SAMPLES OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: LARGE SAMPLE RESIDUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH BD FACSLYRIC¿ FLOW CYTOMETER CARRYOVER INVOLVING PATIENT SAMPLES OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: LARGE SAMPLE RESIDUE.