FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE

MDR report key: 22388868 · Received July 2, 2025

Report

Report Number
9610773-2025-04277
Event Type
Malfunction
Date Received
July 2, 2025
Date of Event
June 16, 2025
Report Date
December 3, 2025
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761075015
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: D8, H2, H3, H6, H11. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: THE COMMUNICATION ERROR B30 WAS DUE TO A COMPONENT FAILURE OF THE UNIVERSAL CABLE. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SUBJECT DEVICE HAS SCOPE COMMUNICATION ERROR B30. THE ISSUE OCCURRED SERVICE / MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
142390 VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE RIGID VIDEO LAPAROSCOPE HET OLYMPUS WINTER & IBE GMBH WA50052A 04042761075015

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown