FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 22384759 · Received July 2, 2025

Report

Report Number
9617229-2025-11067
Event Type
Injury
Date Received
July 2, 2025
Date of Event
June 4, 2025
Report Date
August 13, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
UDI-DI
10888628000322
PMA / PMN Number
P990074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCE'S NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D9, H3, H6. DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS OF DEFLATION WAS RECEIVED ON AUGUST 07, 2025, WITH LOT NUMBER 1202952. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINT CODES ARE: * DEFLATION: OBSERVED AN OPENING ASSESSED AS AN UNIDENTIFIED (TEAR) OPENING. AS PER THE INVESTIGATION PROCEDURE, CREASES WERE OBSERVED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B5 B6 D6B H6.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "DEFLATION". DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

Description of Event or Problem · 0

"HOLE NON-SPECIFIC" OBSERVED DURING SURGERY. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
9027 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 1202952 10888628000322

Patients

Seq Age Sex Outcome Treatment
1 54 YR Female Required Intervention