FDA Adverse Event Injury Summary report: N

AEQ REV II GLENOID BASEPLATE DIA 25MM

MDR report key: 22373461 · Received July 1, 2025

Report

Report Number
3000931034-2025-00341
Event Type
Injury
Date Received
July 1, 2025
Date of Event
June 2, 2025
Report Date
August 29, 2025
Manufacturer
TORNIER S.A.S.
Product Code
PHX
UDI-DI
03700386935800
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SUBJECT DEVICE IS NOT CLEARED FOR SALE IN THE U.S., BUT A SIMILAR DEVICE (K050316) IS COMMERCIALLY AVAILABLE IN THE U.S.

Additional Manufacturer Narrative · 0

THE REPORTED EVENT COULD NOT BE CONFIRMED SINCE THE DEVICE WAS NOT RETURNED FOR EVALUATION AND NO OTHER EVIDENCE WERE PROVIDED. A DEVICE INSPECTION WAS NOT POSSIBLE SINCE THE AFFECTED DEVICE WAS NOT RETURNED FOR INVESTIGATION. FROM THE PROVIDED PICTURE TAKEN JUST AFTER THE DEVICES EXPLANTATION (SOILED IMPLANTS), NO ADDITIONAL INFORMATION COULD BE OBSERVED. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO INDICATIONS OF MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION. MORE DETAILED INFORMATION ABOUT THE COMPLAINT EVENT MUST BE AVAILABLE IN ORDER TO DETERMINE THE ROOT CAUSE OF THE COMPLAINT EVENT. IF ANY ADDITIONAL INFORMATION IS PROVIDED, THE INVESTIGATION WILL BE REASSESSED.

Description of Event or Problem · 0

IT WAS REPORTED THAT, A SHOULDER REVISION WAS PLANNED ON (B)(6) 2025 FOR INFECTION. NO FURTHER INFORMATION WAS AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT, A SHOULDER REVISION WAS PLANNED ON (B)(6) 2025 FOR INFECTION. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1193937 AEQ REV II GLENOID BASEPLATE DIA 25MM SHOULDER PROSTHESIS, REVERSE CONFIGURATION PHX TORNIER S.A.S. AB8259 03700386935800

Patients

Seq Age Sex Outcome Treatment
1 83 YR Female Required Intervention