FDA Adverse Event Malfunction Summary report: N

BD BBL¿ INDIA INK REAGENT DROPPERS

MDR report key: 22373448 · Received July 1, 2025

Report

Report Number
1119779-2025-00766
Event Type
Malfunction
Date Received
July 1, 2025
Date of Event
May 27, 2025
Report Date
July 9, 2025
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
JTS
UDI-DI
30382902611944
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE MANUFACTURING LOCATION FOR THIS PRODUCT IS RR DONNELLEY. THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, BD CORPORATE HEADQUARTERS IN SPARKS, MD HAS BEEN LISTED IN SECTIONS D.3. AND G.1. AND THE SPARKS FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY - BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, RETENTION SAMPLES WERE VISUALLY EXAMINED, AND NO ISSUES WERE OBSERVED RELATING TO PRECIPITATE AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE OF PRECIPITATE. A ROOT CAUSE WAS UNABLE TO BE IDENTIFIED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

REPORT 5 OF 8: IT WAS REPORTED WHILE USING BD BBL¿ INDIA INK REAGENT DROPPERS THAT ONE DROPPER HAD GRAINY DYE MAKING IT UNUSABLE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

REPORT 5 OF 8: IT WAS REPORTED WHILE USING BD BBL¿ INDIA INK REAGENT DROPPERS THAT ONE DROPPER HAD GRAINY DYE MAKING IT UNUSABLE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1357612 BD BBL¿ INDIA INK REAGENT DROPPERS STAINS, MICROBIOLOGIC JTS BECTON DICKINSON & CO. (SPARKS) B03H029M 30382902611944

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown