FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2236866 · Received September 7, 2011

Report

Report Number
2531779-2011-06575
Event Type
Malfunction
Date Received
September 7, 2011
Report Date
April 28, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION COMPLETED ON (B)(4) 2011. RECALL 2531779-03/24/2010-003-R. DURING EVALUATION THE CONTAMINATION WAS FOUND ON THE FORCE SENSOR AND THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION.

Description of Event or Problem · 1

EVALUATION REVEALED CONTAMINATION ON THE FORCE SENSOR ASSEMBLY. THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 27 YR