Narratives
ALTHOUGH THERE WAS NO REPORTED INJURY IN THIS CASE, THE AVAILABLE INFORMATION SUGGESTS A POSSIBLE MALFUNCTION OF THE DEVICE. THOUGH STILL UNDER INVESTIGATION, VARIAN HAS DETERMINED THAT A MDR IS APPROPRIATE AS THIS POSSIBLE MALFUNCTION, SHOULD IT RECUR, COULD POTENTIALLY CAUSE SERIOUS INJURY. ADDITIONAL FOLLOW-UP TO THIS MDR IS EXPECTED UPON COMPLETION OF THE INVESTIGATION.
SITE REPORTS SEVERAL MLC GAP ADJUSTMENTS IN THE TRUEBEAM. THE CUSTOMER REPORTS THAT WHEN USING ENERGIES OF 6FFF AND 10FFF, THE BEAM MODELING IS OKAY - NO PROBLEMS. WHEN THEY PERFORM QA FOR SMALL FIELD SIZES (LESS THAN 12 X 12 CM), THEY NEED A LARGE DOSIMETRIC LEAF GAP (0.25CM). FOR BIG FIELDS, SUCH AS PELVIC GYN AND HEAD AND NECK, THE QAS PASS WITH SMALLER DLG (0.14). THE CUSTOMER REPORTS THAT THEY HAD SEVERAL GAP ADJUSTMENTS IN THE TRUEBEAM. THE FIRST ONE WAS MADE BY THE ENGINEER INSTALLER, BECAUSE THEY WERE HAVING BEAM OFF, DUE THE MLC FAULTS AND SINCE THEN, THEY NEVER GET A VERY GOOD AGREEMENT BETWEEN ECLIPSE AND IMRT QA IN TRUEBEAM. INITIALLY, THEIR PICKET FENCE TEST WAS SHOWING THAT THE MLC WAS SKEWED, THE ENGINEERS THEN ADJUSTED THE MLC WHICH IMPROVED THE PICKET FENCE RESULTS. AT THIS POINT, THEIR SMLC DELIVERY (PLANNED IN PINNACLE) SHOWS GOOD AGREEMENT WITH THE PLANNING SYSTEM AND TRUEBEAM. THEY COMPARE SMLC CALCULATIONS FROM PINNACLE AND ECLIPSE, WHICH ALSO SHOWS GOOD AGREEMENT. THE ONLY PROBLEM IS DMLC DELIVERY ON TRUEBEAM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRILOGY MX | ACCELERATOR, LINEAR, MEDICAL | IYE | VARIAN MEDICAL SYSTEMS, INC | H19 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |