FDA Adverse Event Death Summary report: N

RELIANT STENT GRAFT BALLOON CATHETER

MDR report key: 2234286 · Received August 29, 2011

Report

Report Number
2953200-2011-01488
Event Type
Death
Date Received
August 29, 2011
Date of Event
August 1, 2011
Report Date
August 1, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR SANTA ROSA
Product Code
DQY
PMA / PMN Number
K050038
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL, RESULTS: (DEATH, RUPTURED ANEURYSM/VESSEL), (SEVERELY FRAIL AND THIN AORTIC WALLS), (BALLOON OUTSIDE THE STENT GRAFT). EVAL, CONCLUSIONS: (SEVERELY FRAIL AND THIN AORTIC WALLS), (BALLOON OUTSIDE THE STENT GRAFT).

Description of Event or Problem · 1

AN ENDURANT STENT GRAFT SYS WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5.5 CM IN DIAMETER ABDOMINAL AORTIC ANEURYSM. THE AORTIC NECK WAS 8 MM IN LENGTH AND 21 MM TO 30 MM IN DIAMETER. ILIAC ARTERIES HAD MILD CALCIFICATION AND MODERATE TORTUOSITY. IT WAS REPORTED THAT AFTER THE ENDURANT BIFURCATED STENT GRAFT WAS DEPLOYED, A DISTAL TYPE I ENDOLEAK WAS EVIDENT (MFR REPORT # 2953200-2011-01486). THE PHYSICIAN ELECTED TO INTERVENE BY IMPLANTING AN ILIAC EXTENSION LIMB (MFR REPORT # 2953200-2011-01487). THE PHYSICIAN ELECTED TO MODEL THE STENT GRAFTS WITH A RELIANT BALLOON (MFR REPORT # 2953200-2011-01488); HOWEVER, WHEN MODELING THE LEFT ILIAC ARTERY, THE BALLOON WAS NOT COMPLETELY WITHIN THE STENT GRAFT, BUT NO ISSUES WERE NOTED AT THAT TIME. AT SOME POINT DURING THE PROCEDURE, THE ANEURYSM SAC RUPTURED. THE PHYSICIAN ELECTED TO INTERVENE BY CONVERTING THE PT TO AN OPEN REPAIR. DURING THE OPEN REPAIR, THE PHYSICIAN NOTED THAT THE PT HAD SEVERLY FRAIL AND THIN AORTIC WALLS. THE PHYSICIAN COULD NOT CONTROL THE BLEEDING DUE TO THE FRAIL VESSELS, AND THE PT EXPIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RELIANT STENT GRAFT BALLOON CATHETER DQY MEDTRONIC CARDIOVASCULAR SANTA ROSA NA UNK

Patients

Seq Age Sex Outcome Treatment
1 78 YR Death| R