BD PYXIS¿ MEDBANK MINI
Report
- Report Number
- 2016493-2025-90823
- Event Type
- Malfunction
- Date Received
- June 26, 2025
- Date of Event
- December 5, 2022
- Report Date
- June 25, 2025
- Manufacturer
- CAREFUSION 303, INC.
- Product Code
- BRY
- UDI-DI
- 10885403512582
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING JANUARY 2022-DECEMBER 5, 2024, UNDER CAPA 10308384. THE LATE SUBMISSION OF THIS REPORT IS JUSTIFIED BY THE THOROUGH AND DETAILED NATURE OF THE RETROSPECTIVE REVIEW PROCESS. THIS PROCESS WAS ESSENTIAL TO ACCURATELY CAPTURE ALL RELEVANT DATA AND ENSURE THE INTEGRITY OF OUR REPORTING. WE HAVE INCLUDED THE CAPA REFERENCE FOR THE RETROSPECTIVE REVIEW IN THE ADDITIONAL MANUFACTURER NARRATIVE SECTION OF THE FDA FORM 3500A, AS RECOMMENDED. A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINTS WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 26-DEC-2019 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE DRAWER FAILURE. THE TECHNICAL SUPPORT SPECIALIST ANALYZED THE ISSUE, AND FOUND THE SITE WAS IN DENTIST LOCATION AND CLOSING CASE, 866-907-0860 IS THE SUPPORT NUMBER FOR DENTAL CLINICS TO RESOLVE THE ISSUE. THE SYSTEM FUNCTIONED AS INTENDED AFTER TROUBLESHOT BY THE TECHNICAL SUPPORT SPECIALIST.
IT WAS REPORTED BY THE CUSTOMER THAT BD PYXIS¿ MEDBANK MINI HAD A DRAWER FAILURE, CABINET FREEZE WHEN ATTEMPTED TO DISPENSE OR RETURN MULTIPLE MEDICATIONS. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1856386 | BD PYXIS¿ MEDBANK MINI | AUTOMATED DISPENSING CABINET | BRY | CAREFUSION 303, INC. | 349 | 10885403512582 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |