COOL PATH 7F, 1304-CP-7-25-M
Report
- Report Number
- 2030404-2011-00243
- Event Type
- Death
- Date Received
- August 19, 2011
- Date of Event
- August 12, 2011
- Report Date
- August 12, 2011
- Manufacturer
- ST JUDE MEDICAL, IRVINE
- Product Code
- OAD
- PMA / PMN Number
- P060019
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE DEVICE WAS NOT RETURNED FOR EVAL. REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THE DEVICE MET MFG REQUIREMENTS PRIOR TO SHIPMENT. THE ROOT CAUSE CLASSIFICATION IS CONSISTENT WITH ANTICIPATED PROCEDURAL COMPLICATIONS AS THIS EVENT IS A KNOWN PHYSIOLOGICAL EFFECT OF THE PROCEDURE AND IS NOTED WITHIN THE IFU.
IT WAS REPORTED THAT DURING A VENTRICULAR TACHYCARDIA (VT) ABLATION PROCEDURE WHILE ABLATING WITH THE COOL PATH CATHETER, A STEAM POP OCCURRED WHICH RESULTED IN CARDIAC TAMPONADE AND SUBSEQUENT DEATH. THE PT SUFFERED ELECTRICAL STORM 2 DAYS PRIOR TO ABLATION PROCEDURE AND RECEIVED 7 SHOCKS AND 900 ATP (ANTITACHYCARDIA PACING) THERAPIES FROM HER ICD. THE PT WAS REPORTEDLY VERY UNSTABLE BEFORE THE ABLATION PROCEDURE REQUITING INTUBATION. THE ABLATION PROCEDURE WAS CONSIDERED A "LAST CHANCE" TREATMENT. THE PHYSICIAN WAS ABLATING WITH THE THERAPY COOL PATH ABLATION CATHETER IN THE RIGHT VENTRICULAR OUTFLOW TRACT AND COMPLETED TWO 60 SECOND RADIO FREQUENCY (RF) ABLATIONS. VELOCITY MAPPING, PACE MAPPING AND ANOTHER 60 SECOND RF ABLATION WERE PERFORMED WHEN A STEAM POP WAS HEARD AND CARDIAC TAMPONADE WAS CONFIRMED. THE ABLATION PARAMETERS WERE 30W, 40 DEGREES C, 30 ML/ML WITH A NON-SJM GENERATOR. A PERICARDIALCENTISIS WAS PERFORMED, HOWEVER, NO BLOOD COULD BE DRAINED. THE PT REQUIRED CARDIAC SURGERY TO OPEN THE PERICARDIUM AND MANUALLY REMOVE LARGE BLOOD CLOTS. THE PT SUBSEQUENTLY DIED A FEW MINUTES LATER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COOL PATH 7F, 1304-CP-7-25-M | NONE | OAD | ST JUDE MEDICAL, IRVINE | 83352 | 3265698 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Death |