FDA Adverse Event Death Summary report: N

THERMOSUIT SYSTEM

MDR report key: 2234191 · Received August 31, 2011

Report

Report Number
3006059109-2011-00001
Event Type
Death
Date Received
August 31, 2011
Date of Event
August 1, 2011
Manufacturer
LIFE RECOVERY SYSTEMS HD, LLC
Product Code
NZE
PMA / PMN Number
K061023
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS MANUFACTURED FOR LIFE RECOVERY SYSTEMS BY SUBCONTRACT MFR, NEXCORE TECHNOLOGY, INC., (B)(4). DEVICE EVAL CONDUCTED BY NEXCORE.

Description of Event or Problem · 1

ON (B)(6) 2011, THE PT WAS BROUGHT TO AN OUTLYING HOSPITAL UNRESPONSIVE, APNEIC AND PULSELESS AFTER AN EXTENDED PERIOD OF CARDIAC ARREST. PT WAS TRANSPORTED BY SPOUSE. AFTER 40 MINS OF CPR THE PT RETURNED TO SPONTANEOUS CIRCULATION. THE PT WAS TRANSFERRED TO ANOTHER HOSPITAL VIA AIRLIFT AND WAS TAKEN DIRECTLY TO THE CATH LAB FOR SUCCESSFUL PCI. POST-RESUSCITATION THERAPEUTIC HYPOTHERMIA PER AHA GUIDELINES WAS PRESCRIBED. THE PT WAS POSTURING PRIOR TO COOLING. COOLING WAS INITIATED WITH THE THERMOSUIT SYSTEM WITH A TARGET PT TEMPERATURE OF 33 DEGREES C. THE PT WAS COOLED FOR APPROX 80 MINS, DURING WHICH THE PT TEMPERATURE DID NOT CHANGE ON THE THERMOSUIT SYSTEM. IT WAS DETERMINED THAT THE THERMOSUIT SYSTEM HAD MALFUNCTIONED AND DISPLAYED AN INACCURATE PT TEMPERATURE. ALTHOUGH THE THERMOSUIT SYSTEM DISPLAYED 39 DEGREES C, THE ACTUAL PT TEMPERATURE USING A DIFFERENT MONITOR WAS 28.5 DEGREES C. ACTIVE REWARMING WAS INITIATED. THE PT WENT INTO CARDIAC ARREST, THEN REC'D CPR, HOWEVER WAS UNRESPONSIVE. THE HOSPITAL PHYSICIANS INDICATED THAT THE CAUSE OF DEATH WAS THE UNDERLYING DISEASE AND PROLONGED PERIOD OF CARDIAC ARREST RATHER THAN THE PT OVER COOLING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMOSUIT SYSTEM THERMAL REGULATING SYSTEM NZE LIFE RECOVERY SYSTEMS HD, LLC TSP-50 NA

Patients

Seq Age Sex Outcome Treatment
1 61 YR Death